CAUTION
Disinfectant
Formaldehyde
Peracetic acid
Sodium azide
NaOH
Installation and Operating Instructions Sartocon
Disinfecting the Filter Cassettes
4
Disinfecting the
Filter Cassettes
Sartocon
Slice Disposable has been designed as
®
a ready-to-use product for a single use only.
You must clean and flush the filter cassettes
before disinfecting them. After disinfection,
substances on the membranes can no longer
be removed.
We recommend the following disinfectants and
conditions for the various membrane materials:
Concentration
2–3%
0.1%
0.1%
1 M
Adjust the following pressures:
p
= 29 psi (2.0 bar or 200 kPa)
inlet
p
= 7.3 psi (0.5 bar or 50 kPa)
outlet
p
= 0 psi (0 bar or 0 kPa)
per
— Recirculate the disinfectant through
the cassette(s).
Time
Temperature
30 min
20–30°C
15 min
10–40°C
15 min
20–30°C
30 min
40–50°C
Slice Disposable
®
19