Herunterladen Inhalt Inhalt Diese Seite drucken

Checking The Functionality Of The Filter Cassettes; Manual Integrity Test - Sartorius Sartocon Slice Disposable Installation Und Betriebsanleitung

Filtrationssystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Checking the Functionality of the Filter Cassettes
3
Checking the Functionality
of the Filter Cassettes
Check whether the filter cassettes function properly by measuring their air
diffusion rate. The total air diffusion must not exceed the limit value of a single
filter cassette. In this way, you can ensure that the cassette is intact and will
function correctly.
Make sure to wet the filter cassette before conducting the test
(p
= 43.5 psi (3.0 bar or 200 kPa),
inlet
p
= 14.5 psi (1.0 bar or 100 kPa) for 5 min.)
outlet
3.1

Manual Integrity Test

Proceed as follows:
— Attach a hose to the feed inlet, and connect the other end of the hose
to a compressed air source.
— Close the retentate outlet.
— Adjust the inlet test pressure p
Data"), a keep this pressure constant.
— Close the lower permeate outlet (see Fig. 4).
— At a constant inlet pressure pinlet, wait four minutes until the system
has stabilized.
— Use a flow meter or an inverted burette filled with water to measure
the air diffusion at the upper permeate outlet.
14
Installation and Operating Instructions Sartocon
to the value specified (see "6 Technical
inlet
Slice Disposable
®

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis