Pulse rate measurement
Power supply
External DC power supply (isolated)
Internal battery
Housing construction
Environmental conditions
Electromagnetic compatibility
Potential impact of electromagnetic
disturbances
Air filter
Air tubing
IEC 60601-1 classifications
Range: 40 bpm to 240 bpm, Resolution: 1 bpm, Accuracy: Refer to Nonin Xpod technical
specifications at www.nonin.com.
AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
24 V, 3 A
Lithium-Ion battery, 14.4 V, 2.75 Ah, 40 Wh
Operating hours: 3 h with a new battery under normal conditions (see below).
Patient type: home chronic; pressure: IPAP/EPAP 15/5 cm H
tubing: 2 m; leak: 0; respiratory rate: 20 bpm; battery capacity: 100%
Patient type: hospital acute; pressure: IPAP/EPAP 20/5 cm H
tubing: 2 m; leak: 0; respiratory rate: 45 bpm; battery capacity: 100%
Flame retardant engineering thermoplastic
Operating temperature: 0°C to 35°C
Operating humidity: 10%–95% non-condensing
Storage and transport temperature: -20°C to 60°C (+50°C*)
Storage and transport humidity: 10%–95% non-condensing
Air pressure: 680hPa to 1,100hPa; Altitude: 3,000 m
Note: The performance may be limited below 800 hPa.
*NONIN XPOD
The device complies with all applicable electromagnetic disturbances requirements (EMD)
according to IEC 60601-1-2:2014, for residential, commercial and light industry
environments.
It is recommended that mobile communication devices are kept at least 0.3 m away from
the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device
can be found in www.resmed.com/downloads/devices.
The loss or degradation of the following clinical functions due to electromagnetic
disturbances could result in compromised patient safety:
Accuracy of ventilation control
Accuracy of monitoring of airway pressure and volume
Therapy alarms
Detection of this degradation could be observed by the following device behaviour:
Erratic ventilation delivery
Rapid fluctuations in monitored parameters
False activation of therapy or technical alarms
Electrostatic fibre mesh with TPE frame structure. Bacterial filtration efficiency of >99.9%
for ~2.7 micron particle size.
Standard air tubing: Flexible plastics, 2 m or 3 m length (22 mm diameter)
SlimLine air tubing: Flexible plastics, 1.83 m length (15 mm diameter)
For more information on air tubing resistance to flow, compliance, tubing length, and inner
diameter, refer to the Air tubing compliance guide in ResMed.com.
Class II (Clause 3.14–double insulation). This adherence means the need for an
protective earthing (ie, an earthed plug) is not necessary.
Type BF
Continuous operation
O; mask type: Ultra Mirage; air
2
O; mask type: Ultra Mirage; air
2
English
31