Installation
The display can be turned in stages of 90°. First remove the cover clamp (1) and the housing
cover (2). Then remove the display (3) from the electronics unit (4). For configuration purposes,
the ribbon cable between the display and electronics unit has to be plugged in.
Turn the display to the desired position and then attach it to the electronics unit.

3.3.2 Direct wall mounting

Proceed as follows to mount the device directly on the wall:
• Drill 2 holes
• Attach the device to the wall with 2 screws (∅5 mm (0,2 in)).
3.3.3 Pipe mounting
The mounting bracket is suited for pipes with a diameter between 38 to 84 mm (1.5 to 3.3 in).
Proceed as follows to mount the device on a pipe:
• Attach the mounting bracket to the pipe
• The additional mounting plate must be used for pipes with a diameter of 38 to 56 mm (1.5 to
2.2 in).
• Fix the device to the mounting bracket with the two screws supplied. The mounting plate is
not needed for pipes with a diameter of 56 to 84 mm (2.2 to 3.3 in).
Fig. 5: Mounting the field indicator on a pipe with mounting bracket for pipe diameter 1.5 to 2.2 in
Mounting kit, consisting of:
1:
Mounting plate
2:
Mounting bracket
3:
2 nuts M6