Herunterladen Inhalt Inhalt Diese Seite drucken

wissner-bosserhoff SafeSense Gebrauchsanweisung Seite 54

Bed-exit system
Vorschau ausblenden Andere Handbücher für SafeSense:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Table 3: Guidelines and manufacturer's declaration - electromagnetic immunity - for all ME equip-
ment and ME systems that are not life-sustaining
SafeSense
is designed to operate in an electromagnetic environment as specified below. The customer
®
or user of SafeSense
Interference im-
IEC 60601-
munity checks
test level
3 Vrms
150kHz bis 80MHz
Conducted RF
(IEC 61000-4-6)
6 Vrms
150kHz bis 80MHz
3 V/m
80MHz bis 2,7GHz
Radiated RF
(IEC 61000-4-3)
10 V/m
80MHz bis 2,7GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
54
should ensure that it is operated in such an environment.
®
Compliance level
3 Vrms
6 Vrms
3 V/m
10 V/m
EMC guidelines
Portable and mobile communica-
tion devices should not be used
when any closer to SafeSense
including the cables, than the
recommended separation distance
calculated according to the applica-
ble equation for the transmission
frequency.
Recommended separation distance
d=0,35*√P
80 MHz bis 800 MHz
d=0,35*√P
d=0,35*√P
800 MHz bis 2,7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range.
b
Interference may occur in the vici-
nity of equipment marked with the
following symbol.
,
®
, should
a

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis