wissner-bosserhoff SafeSense Gebrauchsanweisung Seite 53

Bed-exit system
Vorschau ausblenden Andere Handbücher für SafeSense:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Table 2: Guidelines and manufacturer's declaration - electromagnetic immunity - for all
ME equipment and ME systems
SafeSense
is designed to operate in an electromagnetic environment as specified below.
®
Interference immunity
checks
Electrostatic discharge
(ESD)
(IEC 61000-4-2)
Electrical fast transi-
ent¬/burst
(IEC 61000-4-4)
Surge
(IEC 61000-4-5)
Voltage dips, short in-
terruptions and voltage
variations on power
supply lines
(IEC 61000-4-11)
Magnetic field power
frequency (50/60 Hz)
(IEC 61000-4-8)
NOTE: U
is the AC supply voltage prior to application of the test level
T
IEC 60601 test
Compliance level
level
± 8 kV contact
± 8 kV contact
discharge
discharge
± 15 kV air
± 15 kV air
discharge
discharge
± 2 kV for power
± 2 kV for power
cables
cables
± 1 kV for input
± 1 kV for input
and output lines
and output lines
± 1 kV voltage
± 1 kV voltage
line-line
line-line
± 2kV voltage
± 2 kV voltage
line-earth
line-earth
< 5% U
(> 95%
< 5% U
T
dip in U
) for ½
dip in U
T
cycle
cycle
40% U
(60% dip
40% U
T
in U
) for 5 cycles
in U
T
70% U
(30% dip
70% U
T
in U
) for 25 cycles
dip in U
T
cycles
< 5% U
(95% dip
T
in U
) for 5 sec.
< 5% U
T
dip in U
sec
30 A/m
30 A/m
EMC guidelines
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%
Grid power quality should be
that of a typical commercial or
hospital environment
Grid power quality should be
that of a typical commercial or
hospital environment.
(> 95%
Grid power quality should be
T
) for ½
that of a typical commercial or
T
hospital environment.
(60% dip
T
) for 5 cycles
T
If the user requires SafeSense
(30%
to continue to function even in
T
) for 25
the event of interruptions to the
T
power supply, we recommend
connecting the Bed-Exit System
(>95%
to an uninterruptible power
T
) for 5
supply or a battery.
T
The magnetic field power
frequency should be at levels
characteristic for a commercial
or hospital environment.
®
53