Perpendicular adjustment of the castor bearing blocks
To achieve an optimum performance, the bearing blocks must always be
aligned vertically to the ground. Some Adjustments result in a change of the
seat angle. the vertical position of the bearing blocks should then be
GB
adjusted.
Method
Slacken the two screws on the bearing block (10mm open-ended spanner,
steadying with 5mm Allen key). Rotate the hexagonal/octogonal cam to
adjust the bearing block vertically and re-tighten.
CAUTION:
angle d'inclinaison important
Octogonal
Excentrique
cam
à 8 pans
ULTRA-LIGHT NEO with titanium or Carbon front frame
Since the bearing blocks are fixed, no adjustment is possible.
Therefore, ensure that the seat angle of the wheelchair is always 4.2
+/- 1.5cm (seat angle = seat height at front - seat height at back).
34
Ensure that the octogonal cam adjustments on both
sides of the wheelchair are identical (i.e. when the
adjustments are the same, viewed from the fitting
side, on one side the stamp marking on the octogonal
cam is visible and on the opposite side it is not).
(not valid for the hexagonal cam)
Large seat angle
angle d'inclinaison rédu
Small seat angle
Excentrique
Octogonal
à 8 pans
cam
Fourche