D D i i s s i i n n f f e e c c t t i i n n g g
W W A A R R N N I I N N G G
• Always unplug the product before you begin cleaning or disinfecting.
• Do not immerse the product in cleaning or disinfectant solutions.
• Do not spray or allow liquids to enter the power cord ports or the hose ports.
• Always make sure that you wipe each product with clean water and dry each product after you clean. Some cleaning
agents are corrosive in nature and may cause damage to the product. Failure to follow the cleaning and disinfecting
instructions may void your warranty.
C C A A U U T T I I O O N N
• Do not over expose the product to higher concentration chemical solutions as these may degrade the product.
• Failure to follow manufacturing instructions may also affect useful life of the product.
Recommended chemical solutions:
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic solution (Matar)
Chlorinated bleach solution (6500 ppm)
Accelerated hydrogen peroxide
Potassium peroxymonosulfate
70% isopropyl alcohol
Follow hospital protocol for support surface care between patients to avoid the risk of cross-contamination and infection.
N N o o t t e e - - Make sure you unplug the product before you begin cleaning and disinfecting.
EN
20
2874-009-005 Rev A.0