Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
1. On/Off key
2. Set/Menu:
3. Tare:
4. Store/Quit: Activate hold function or logger function (see Section 3)
5. Min/Max:
Max-Function: Pressing ´Max´ shows the maximum measured value
Min-Function: Pressing ´Min´ shows the minimum measured value
Hold-Function: By pressing ´Store/Quit´ the instantaneous measured
Logger-Functions: Activated by the ´Store/Quit´ key, only if the logger
Tare-Functions: By pressing ´Tare´, the display will be set to 0. All meas-
Activating/deactivating tare clears the max- & min-memories.
Zero-Point Adjust: (for gauge pressure sensors only) If there is no pres-
A zero-point adjustment can only be carried out if the difference between
the value on display is less than 500 digits!
If a zero-point adjustment has been made, this will be signalled by the
message "nuLL-Corr" being displayed briefly as the device is turned on.
Differential pressure: Pressing ´Set/Menu´ shows the difference between
WIKA Operating Instructions Hand-Held Pressure Indicator
Enter Configuration Menu (press for 2 sec)
Activate tare function, zero point adjustment
Display the respective min-/max-memory value in lower
display
recorded. Pressing it again hides it. To clear the max-
memory press the ´Max´ key for > 2 seconds.
recorded. Pressing it again hides it. To clear the min-
memory press the ´Min´ key for > 2 seconds.
value will be shown in the lower display. Pressing it again
hides it. (Only when the main menu item logger = 'off' is
selected).
has been selected via the main menu (see Section 3).
urements from then on will be displayed relative to the set
tare value. When the tare function is activated, the arrow
´Tare´ appears in the display. To deactivate the tare func-
tion press ´Tare´ for > 2 seconds.
sure applied to the pressure ports the device will display 0.
If there is a permanent offset (and the device is being used
under steady conditions), a permanent zero point adjust-
ment can be carried out. To carry out the adjustment,
press ´Tare´ for approx. 5 seconds. (Please note: A zero-
point adjustment can only be made if the difference bet-
ween the display value and the factory calibration value is
less than 2 %! E.g. for a measuring range of 0 ... 25 bar,
=> zeropoint adjustment up to 0.5 bar possible). To revert
to the factory calibration, press ´Tare´ for approx. 15 sec-
onds.
the measured values of Channel 1 and Channel 2 (CH1-
CH2) in the lower display. Pressing it again will hide it (only
with the 2-channel version and when 2 sensors are con-
nected).