Specification
Measurement
Method
Light source
Excitation wavelength
Detection wavelength
Measurement range
Display units
Measurement interval
Event log and data log
Accuracy
Limit of detection
Response time
Turbidity compensation
Calibration verification
Calibration
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

2.1 Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
4 English
Details
UV fluorescent measurement process for poly-cyclic aromatic hydrocarbons
(PAH)
Ultraviolet light emitting diode
255 nm
370 nm
0 to 900 μg/L PAH
(phenanthrene) equivalents
PHE
ppb, ppm, μg/L, mg/L PAH
10 seconds
2 weeks capacity minimum, all measurements
± 5% or ± 5 μg/L (the larger value) at a constant temperature and flow
3 μg/L PAH
10 seconds
0 to 150 FNU (or 0 to 150 NTU); SC1000 Controller or other programmable
controller calculates turbidity compensation based on the PAH reading and
turbidity reading.
Note: The user must manually enter the turbidity compensation formula into the controller.
Refer to
Figure 8
on page 15 for the formula.
Calibration standard measured
Factory calibration with Phenanthrene equivalent standards
Optional: 1-point or 2-point user calibration with the Phenanthrene equivalent
standards in the Calibration Kit.
As an alternative, the calibration slope and offset values can be changed
manually.