6 Troubleshooting
Malfunction
Device does
not start.
Device does
not charge.
Insufficient
suction
No suction
Device swit-
ches off.
Contact ASSKEA GmbH or your service partner if a
malfunction cannot be corrected by the described measures.
Display
Probable cause
do not illuminate
Battery is discharged
blinks red
Battery is discharged and impermissible
input voltage
blinks red
Internal error
blinks red
•
Battery temperature too
high for charging
and
illuminates green
•
Battery is defective.
blinks red
Impermissible input voltage
illuminates green /
•
Leak in the suction line or in the canister
orange
and / or
illuminates green
•
Service life of vacuum generator expired
continuously
•
Float ball of the overflow protection is
illuminate red
blocking the aspiration lid inlet (reusable
secretion canister)
•
"OneWay" disposable liner is full.
•
Bacterial filter is blocked.
•
Suction tube is blocked.
•
Adapter for the disposable secretion
canister system or bacterial filter is not
inserted.
blinks red
Internal error
4 Maintenance
Remedy
Connect the power supply unit
for charging or operating.
Connect the correct power supply
unit, check the power supply.
Please contact service!
•
No action required, automatic
resumption of charging when
battery has cooled down
•
Please contact service!
Connect the correct power supply
unit, check the power supply.
•
Check that the lid of the "OneWay" dis-
posable liner or the reusable secretion
canister lid and tube connections are
tight and no leakages are apparent.
•
Please contact service!
•
Check fluid level in the reusable
secretion canister, empty the reusable
secretion canister.
•
Replace the "OneWay" disposable liner.
•
Replace the "OneWay" disposable
liner or the bacterial filter.
•
Replace the suction tube.
•
Insert the adapter for the disposable
secretion canister system or the
bacterial filter.
Please contact service!