4.3.4 Replacement of the "OneWay" disposable liner

Health risks due to handling of infectious
or pathogenic germs
Infectious and pathogenic germs in the aspirate
cause health risks.
Wear disposable gloves when replacing the
"OneWay" disposable liner.
Use the "OneWay" disposable liner for one patient only.
Replace the "OneWay" disposable liner incl. the suction
tube in accordance with the applicable hygiene instruc-
tions if it is full, prior to each new patient, or weekly at the
latest.
For each new patient, professional hygienic processing by
ASSKEA GmbH or an authorized service partner of ASSKEA
GmbH is strictly required!
After each aspiration procedure, all components that
have come into contact with the aspirate must be cleaned,
disinfected, or disposed of.
The disposal of aspirate and contaminated components
must be performed in a proper manner.
1. Switch off the ASSKEA proVisio
holding the On / Off button (Fig. 1.2(C)) pressed for 1-2 seconds.
2. Separate the suction tube (Fig. 1.2(G)) from the patient connection
(Fig. 11(B)).
3. Remove the "Bag" external canister incl. the "OneWay" disposable
liner from the device.
4. Separate the pre-assembled connecting tube of the "OneWay"
disposable liner at the upper end of the T-piece of the "Bag"
external canister.
M28 secretion aspirator by
®
5. Remove the "OneWay" disposable liner from the reusable
"Bag" external canister.
6. Dispose of the "OneWay" disposable liner incl. suction tube
in a proper manner (see section 7.3 "Disposal").
7. Insert a new "OneWay" disposable liner in the reusable "Bag"
external canister as specified in section 4.2.2. Ensure that the
bottom end of the T-piece is fixed tightly in the holder for the
"Bag" external canister as well as the lid of the disposable liner
has to be seated properly on the external canister.
8. Attach a new suction tube to the patient connection
(Fig. 11(B)) of the "OneWay" disposable liner.
9. Switch on the ASSKEA proVisio
by pressing the On / Off button (Fig. 1.2(C)).
4 Maintenance
M28 secretion aspirator
®
EN
29