Hazard to persons due to improper handling
Please read through sections 4.1 and 4.2 carefully!
Perform aspiration in the respiratory tract only after
instruction by specialized personnel!
Use only suitable sterile disposable catheters for
aspiration! Commercially available sterile disposable
catheters from CH10 to CH16 that are intended for
use in the tracheal region can be connected.
If aspiration procedures are carried out too often,
slight bleeding may occur.
EN
Malfunction due to aspirated secretion
Change the reusable secretion canister when it is half-
18
full to curtail frothing. If the reusable secretion canister
is full, the mechanical overflow protection is triggered.
This disrupts the aspiration process.
Switch off the device when emptying the reusable secre-
tion canister.
Ensure that the "OneWay" disposable liner is replaced
on regular basis. If the "OneWay" disposable liner is full,
the integrated overflow protection is activated. This dis-
rupts the aspiration process.
Switch off the device when replacing the "OneWay"
disposable liner.
If secretion has been aspirated into the device, it must
be properly processed by ASSKEA GmbH or by a service
partner authorized by ASSKEA GmbH!
Damage to the device due to insufficient acclimatization
After the device has been exposed to temperatures between
-25 °C and +70 °C during transport / storage according to
the technical data (see section 8), it must first acclimatize
for approx. 2 h at room temperature (ca. 20 °C) before it can
be used as intended.

4.1 Setup and startup

The following section describes the operating elements, connections,
and startup of the ASSKEA proVisio
4.1.1 Connecting the ASSKEA proVisio® M28 secretion aspirator
Known or identifiable conditions for medical
care within a domestic environment
Children and pets should be kept away from the device
to ensure that it cannot be knocked over or dropped.
Prior to connecting the power supply unit, ensure that the
voltage of the device corresponds to the domestic power
supply.
The device may not be used in damp rooms, bathrooms,
or showers.
Avoid wet conditions on power supply unit, control panel,
and socket for the power supply unit.
Never submerge the device in water or other liquids
(including while it is switched off).
Check the power supply unit and mains cable for
potential damage before each use and replace it if there
is any damage.
Use the socket for the 12 V power supply unit (Fig. 1.1(D) and 1.2(D))
to connect the device to the mains power supply with the aid of the
supplied power supply unit to charge or operate it as needed. Make
sure that the device is positioned in such a way that it is possible to
easily disconnect it later. First connect the power supply unit to the
power supply socket of the ASSKEA proVisio
aspirator and then to the mains power supply.
M28 secretion aspirator:
®
M28 secretion
®