5.2 Cleaning and care for homecare use
5.2.1 General information
Health risks due to handling of infectious
or pathogenic germs
Infectious and pathogenic germs in the aspirate
cause health risks.
•
Never operate the device without the bacterial filter.
•
Replace the bacterial filter for the reusable secretion
canister system at least every 2 weeks when used
exclusively for one patient, if it is used frequently!
•
Wear suitable disposable gloves while doing so.
•
The reuse of bacterial filters for multiple patients is
prohibited for hygiene and safety reasons!
•
Replace the "OneWay" disposable liner incl. suction tube
weekly at the latest, prior to each new patient, or if it is full.
•
Prior to passing the device on to another patient or to
other persons, professional hygienic cleaning by ASSKEA
GmbH or by an authorized service partner of ASSKEA
GmbH is strictly required! For that purpose observe the
information regarding the further use of the device (see
section 5.3).
•
After each aspiration procedure, all components that have
come into contact with the aspirate must be cleaned,
disinfected, or disposed of.
•
The aspirate and the parts contaminated with aspirate must
be disposed of properly.
Health risks or damage to the device due to
handling of disinfectants
If you have the opportunity to use disinfectants, please
observe the information in section 5.1. Otherwise,
please follow the instructions from section 5.2.2 to 5.2.7!
Possible bodily injury due to electric shock
•
Switch off the device prior to cleaning / disinfection.
•
Disconnect the power supply unit by unplugging it from
the mains supply. Subsequently disconnect the power
supply unit from the power supply socket of the ASSKEA
proVisio
M28 secretion aspirator.
®
5.2.2 Cleaning of the surface of the device
Clean the surfaces of the device regularly.
The device can be wiped with a damp, lint-free cloth. Repeated
cleaning procedures may result in minor discolorations of the plastic
components of the housing. However these do not affect the function
of the device.
If the interior of the device comes into direct contact with
liquids or solids, the device must be inspected by ASSKEA
GmbH or by a service partner authorized by ASSKEA GmbH.
4 Maintenance
EN