6. Fold the moving part of the secretion canister lid (visioClick
upwards to release the hooks and remove the lid from the secre-
tion canister.
Fig. 25
7. Empty and clean the reusable secretion canister and the lid
according to section 5.1.3 and 5.2.3.
8. Remove the used bacterial filter before a new patient uses the
device or when it has already been used frequently for one patient
for 2 weeks. Move the lever that is in the horizontal position
counterclockwise a quarter turn so that the lever points vertically
downwards.
Fig. 26
)
®
Fig. 27
9. When the lever is in this position, pull the bacterial filter
out of the fitting and dispose of the bacterial filter properly.
Fig. 28
10. Insert a new bacterial filter (see section 4.2.1).
11. Clean the surface of the device according to section
5.1.2 and 5.2.2.
12. Store the device in the shipping carton until next operation.
4 Maintenance
EN