Menu
Param. Setting
„Menu"
SEt
Set Alarm: Alarm function configuration
AL.
AL.
[1,2,DIF]
AL.Lo
[1,2,DIF]
AL.Hi
[1,2,DIF]
SEt
Set Logger: Logger function configuration
LoGG
Func
CYCL
Lo.Po
SEt
Set Clock: Setting of real time clock
CLOC
CLOC
dAtE
YEAr
12
Meaning
▲ or ▼
on
Alarm function on, with audio alarm
no. So
Alarm function on without audio alarm
oFF
No alarm function
Min-Range ...
Min alarm limit
... AL.Hi
(not when AL.oFF, Min-Range is the lower display range of the
connected sensor)
AL.Lo ...
Max alarm limit
... Max-Range
(not when AL.oFF, Max-Range is the upper display range of the
connected sensor)
CYCL
Cyclic: logger function 'cyclic logger'
Stor
Store: logger function 'discrete value logger'
oFF
No logger function
1 ... 3600
Cycle time for cyclic logger [seconds]
on / oFF
Low-power logger with very low power consumption (only if
cyclic logger with slow measuring rate)
HH:MM
Clock: set time hours:minutes
TT.MM
Date: set time day.month
YYYY
Year: set time year
* If there is existing data in the logger memory, the menus/parameters
cannot be accessed! If these need to be adjusted, the logger memory
must first be cleared! (see Section "3. Operation of logger".)
** This menu can oly be invoked if a refering sensor is connected to
connection 1. When using a second sensor at connection 2 then
changes are taken over.

2. Configuration

To change device settings, press the ´Set/Menu´ key for 2 seconds. This
will access the configuration menu (main display: "SEt").
Pressing the ´Set/Menu´ key again scrolls down the menus, pressing
the ´Tare´ key scrolls across to the associated parameters, which can
then be selected with the ´Tare´ key.
The parameters can be altered using the ´Min´ or ´Max´ keys. Pressing
the ´Set/Menu´ key again saves the settings and returns you to the main
configuration menu.
The ´Store/Quit´ key exits the configuration and returns you to standard
measuring mode.