Herunterladen Inhalt Inhalt Diese Seite drucken

Marzhauser Sensotech SensorReady 3D Betriebsanleitung Seite 72

Usb-messinterface für inkrementelle längenmesssysteme
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Putting into Service
NOTICE
Equipment damage possible by connecting inappropriate
measuring systems.
Only connect measuring systems from Märzhäuser Wetzlar to
port Encoder 1.
5 Connect your measuring systems 1 and 2 to port Encoder 1
WARNING
Possible electric shock and equipment damage by connec-
tion to external or damaged power supplies.
Only use the 24 V/30 W power supply provided. Immediately
replace the power supply if the power supply housing or cable
show signs of damage. Only use replacement components from
Märzhäuser Sensotech.
Possible electric shock and equipment damage by connec-
tion to a non-specified mains.
Ensure that the voltage of your mains matches the label on
the power supply before connecting the power supply to the
mains.
6 Connect the power supply cable to the port 24V DC/2A
7 Connect the power supply to the mains.
The status LED lights up.
8 Move your positioning system at least 1 cm in both directions.
72
on the SensorReady 3D. Use the MR encoder cable for this.
Note: The reference switches for your measuring systems 1
and 2 are thus also connected to the SensorReady 3D.
max.
Note: The measuring system is automatically calibrated
in this way. The calibration is kept after switching off the
SensorReady 3D and is continually updated during operation.
Operating Manual

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis