Herunterladen Inhalt Inhalt Diese Seite drucken

Dräger Polytron 6100 EC WL Gebrauchsanweisung Seite 31

Vorschau ausblenden Andere Handbücher für Polytron 6100 EC WL:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Bit
DIAG_STATUS
22
ENVIRONMENTAL
21
FAULT_PREDICTION
20
POWER_CRITICAL
19
POWER_LOW
18
SW update incomplete
17
SIMULATION_ACTIVE
16
– (Bit not used)
15
– (Bit not used)
14
OTHER_FAULTS
13
– (Bit not used)
12
– (Bit not used)
11
– (Bit not used)
10
– (Bit not used)
9
– (Bit not used)
8
– (Bit not used)
7
– (Bit not used)
6
SENSOR_WARMUP
5
SENSOR_CAL_INTERVAL_-
EXPIRED
4
SENSOR_SELFT-
EST_WARNING
3
– (Bit not used)
2
– (Bit not used)
1
– (Bit not used)
0
– (Bit not used)
Dräger Polytron 6100 EC WL, Dräger Polytron Repeater ISA
General
Cause
status
O
Gas detector is operated
outside the specified tem-
perature range.
M
Device warning
F M
Battery charge critical.
M
Battery charge low.
C
Firmware update active
C
Gas detector is in simula-
tion mode.
F
Serious device fault, vari-
ous causes.
O
Warm-up phase 1 active.
M
Calibration interval expired. Calibrate the gas detector.
M
Sensitivity of the sensor is
low.
Troubleshooting
Remedy
Use weather protection kit (accessory)
when exposed to sunlight.
A Check ISA100 Wireless
B Check detail status with PolySoft.
Change battery pack immediately.
Change battery pack soon.
Wait for firmware update (maximum 15
minutes).
Exit simulation mode.
Have the gas detector checked by Dräger-
Service.
Wait for warm-up phase 1 to finish.
Provide for sensor change promptly.
en
connection.

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Polytron repeater isa

Inhaltsverzeichnis