The sensor self-test is a safety feature that is performed at
recurring intervals. If the sensor does not pass a sensor
self-test, a warning is output. Further negative self-tests
will result in an error message being output.
The sensor self-test can be disabled in PolySoft or carried
out manually.
– Gas settings
Depending on the sensor, the gas type, measuring range
and measuring units can be set with this feature.
13.2 Resetting the gas detector to the
factory settings
Can be performed from 10 minutes after turning on the gas
detector.
Data for integration into the ISA100 Wireless
®
Bluetooth
pairing PIN are reset.
Feature is also available in PolySoft.
Procedure using the magnetic pen
Reset factory settings:
1. Hold the magnetic pen on the marking (Fig. A 11) for
approx. 15 seconds until the blue LED flashes rapidly.
2. Withdraw the magnetic pen.
3. Hold the magnetic pen on the marking again within 5 to 10
seconds (the LED stops flashing) and withdraw it.
The gas detector restarts and is reset.
13.3 Changeable safety-relevant factory
settings
PolySoft
Factory setting
SIL status
off
000000
®
Bluetooth
pairingPIN
Sample gas
sensor-specific
Sample gas module
sensor-specific
Measuring range
sensor-specific
Upper capture value
sensor-specific
limit
Lower capture value
sensor-specific
limit
Calibration gas
sensor-specific
Calibration gas module sensor-specific
Concentration of the
sensor-specific
calibration gas
Calibration interval
sensor-specific
A1 alarm
sensor-specific
Dräger Polytron 6100 EC WL, Dräger Polytron Repeater ISA
PolySoft
A2 alarm
14
The gas detector contains lithium-ion batteries. During
transportation, especially when transporting the gas detector
by air, observe the relevant safety regulations and marking for
lithium batteries.
15
network and
Disposal of electrochemical sensors
WARNING
Danger of explosions and risk of chemical burns!
Sensor fluids may leak out and can cause acid burns.
► Do not dispose sensors in fire
► Do not open with force
16

16.1 General data

Housing material
Housing protection class IP 66/67
Weight of gas detector
Weight of battery pack
For information about the dimensions, see appendix
"Dimensions".

16.2 Power supply

Wired power supply
Cable type
Conductor
Thread of cable entry
Operational voltage
Inrush current
Factory setting
sensor-specific

Transport

Disposal
This product must not be disposed of as municipal
waste. It is therefore marked with the adjacent symbol.
This product can be returned to Dräger free of charge.
For information please contact the national sales
organisations and Dräger.
Batteries and rechargeable batteries must not be dis-
posed of as municipal waste. They are therefore
marked with the symbol on the left. Collect batteries
and rechargeable batteries according to local regula-
tions and dispose of at battery collection centres.

Technical data

Stainless steel 316
approx. 3.2 kg (7.0 lb)
approx. 0.5 kg (1.1 lb)
2-core shielded cable
24 to 12 AWG / 0.2 to 2.5 mm²
M 25 x 1.5
10...30 V
2,3 A for 2 ms with 24 V
resistance
Transport
en
, 10 Ω