7. Additional notes for instruments with EHEDG and 3-A

7. Additional notes for instruments with EHEDG and 3-A

7.1 Compliance with 3-A conformity

For a 3-A-compliant connection with milk thread fittings per DIN 11851, suitable profile
sealings must be used (e.g. SKS Komponenten BV or Kieselmann GmbH).

7.2 Compliance with EHEDG conformity

For an EHEDG-compliant connection, sealings in accordance with the current EHEDG
policy document must be used.
Manufacturer of sealings
Sealings for connections per ISO 2852, DIN 32676 and BS 4825 part 3:
e.g. Combifit International B.V.
Sealings for connections per DIN 11851: e.g. Kieselmann GmbH
Neumo BioConnect

7.3 Mounting instructions

Observe the following instructions, especially for EHEDG certified and 3-A-compliant
instruments.
To maintain the EHEDG certification, one of the EHEDG-recommended process
connections must be used.
To maintain the conformity to the 3-A standard, a 3-A-compliant process connection
must be used.
Mount the electrical thermometer, including the thermowell, with minimal dead space
and able to be cleaned easily.
Installation position of the electrical thermometer including thermowell should be
designed to be self-draining.
The mounting position must not form a draining point or cause a basin to be formed.

7.4 Cleaning in place (CIP) cleaning process

When cleaning from outside ("wash down"), observe the permissible temperature and
ingress protection.
Only use cleaning agents which are suitable for the seals used.
Cleaning agents must not be abrasive nor corrosively attack the materials of the wetted
parts.
Avoid thermal shocks or fast changes in the temperature. The temperature difference
between the cleaning agent and rinsing with clear water should be as low as possible.
Negative example: Cleaning with 80 °C and rinsing at +4 °C with cold water.
WIKA operating instructions model TR25 (Ex i)
sealings: e.g. Neumo GmbH & Co. KG
®