Herunterladen Inhalt Inhalt Diese Seite drucken

When To Accept Or Reject An Ear Shell? If A Shell Receives Invalid Results - Phonak SafetyMeter Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6
"Attenuation control" mode
6.8 When to accept or reject an ear shell?
If a shell receives invalid results (
to Phonak Communications for replacement. The SafetyMeter check should be performed
again, with both user's shells, when the replacement shell has been received.
6.9 Certificate description
Peter Smith - Hearing protection service technician
Phonak Communications AG
Laenggasse 17
3280 Murten
Switzerland
SafetyMeter - personal attenuation control certificate
User name:
Product:
Company:
The protection is valid
Maximum recommended exposure to noise when using this hearing protection device:
8 hours
109dBC
* The Personal Attenuation Rating (PAR) is based on the result of a single user test with SafetyMeter.
The PAR is the "single user equivalent" of a Single Number Rating (SNR) as measured per ISO 4869. The PAR is
computed like an SNR except it is calculated individually for the user and does not include a standard deviation
correction.
The [PAR-1.1dB] is the "single user equivalent" of a Noise Reduction Rating (NRR) as measured per ANSI S3.19-1976.
The [PAR-1.1dB] is equal to an NRR calculated individually for the user, i.e. excluding standard deviation corrections
and 3dB spectral uncertainty factor.
In case of doubt, please use the SNR and NRR attenuation values as given in the user instructions.
www.phonak-communications.com
15
004 mode emploi.indd 15
Hans Muster
DWOM Nr:
1234567890
Serenity SP/SPC 100YE
Operator:
Peter Smith
Phonak Communications
Date:
13.05.2010
Left
Right
*
PAR
27 dB
Left
Right
4 hours
2 hours
1 hour
112dBC
115dBC
118dBC
) 3 times in a row, it should be rejected and returned
Operator's header, as defined under
Parameters/Certificate header.
User and test information.
Test result, i.e. attenuation level of the
user's protection.
If the hearing protection is valid, maximum
exposure times are also recommended.
These values only apply to this specific
protection system.
24.6.2010 15:33:40

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis