TROUBLESHOOTING

PROBLEM
1. While pressing the handset button the
LED is not lit.
2. In the case of poor nebulisation
3. If the unit does not nebulise
4. If the appliance interrupts nebulisation
and the green LED flashes, this means
there is no more medication
5. Failure to clean the nebuliser from drug
deposits may endanger proper operation
of the unit.
Should the unit again fail to operate after following the above instructions, please con-
tact your local dealer or authorised after-sales centre.
REMEDY
- When operating through mains sup-
ply make sure that the handset and the
power supply unit are properly plugged
in.
- When operating through a cigarette-
lighter socket (optional accessory),
check that the cord is properly plugged
in and the fuse is not burnt out.
- Make sure that the drug is not too cold
– drug temperature should be close to
room temperature.
- Make sure the drug is not glycol-based
and always remember that the use of
an ultrasonic nebulizer is not recom-
mended.
- Check that the nebulising chamber con-
tains medication.
- Make sure that quantity of medication is
between 4 and 8 ml max.
- Check that the medication is diluted suf-
ficiently with saline solution 0.9%.
- Make sure that the flow adjustment
knob is not in the "STOP" position.
- Check that the transparent cover (A6)
and the disk positioned below are clean.
- Put some more medication into the tray
or proceed to clean and disinfect.
- Carefully follow the instructions for
CLEANING, SANITIZATION AND DISIN-
FECTION.