ATTENTION: When the medication is about to run out, nebulisation becomes intermittent.
Switch off the application by pressing the button (A1). Despite the unit having a sensor
indicating the presence of medication, it is always best not to use the appliance dry
(without medication).
AUTOMATIC SWITCH-OFF: The appliance switches off automatically after about 10
minutes; nonetheless, if you wish to continue with the therapy, simply press the button
again and the appliance will start operating.

CLEANING, SANITIZATION AND DISINFECTION

When treatment is completed remove the power cord from the handset keeping it
away from wet or damp areas. Turn the adjustment knob to the "STOP" position.

CLEANING THE OUTSIDE OF THE DEVICE

Only use a cloth dampened with antibacterial detergent (non-abrasive and free of solvents
of any kind).

CLEANING THE TRANSDUCER

1. Remove the mask and, if necessary, the nasal prong or the
mouthpiece.
2. Pour a spoon (4 ml) of demineralised or drinking water in
the transducer area.
3. Put the respiratory mask back in place.
4. Use the supplied cap to close the nebulisation hole located
inside the mask (as shown in the picture). The cap should be
used only for cleaning purposes.
5. Insert again the power cord and operate the unit for 1-2
minutes.
6. Remove the power cord and the respiratory mask
14