CANISTER HOLDER
For convenience, the canister holder can be detached and reattached on the most suitable side (left or right) as shown
in the figure:
TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
The device does not work
Lack of suction
Lack of suction caused by fluid leakage
Blocked float
Vacuum power poor and/or non-existent
If, after checking the above-described conditions the device is still not working properly, we recommend taking it to your nearest
trusted dealer or authorised FLAEM service centre.
Using a tool, unfasten the canister holder from the bottom of the
device and rotate it in order to release it.
Refasten it, using the inverse procedure, on the selected side.
Ensure that the canister holder is properly fastened to the bot-
tom of the device. This system allows up to two canister holders
to be connected.
CAUSE
The power cable has not been correctly in-
serted into the socket of the device or the
power socket
The collection cannister lid has not been
correctly positioned onto the cannister
Lid gasket not in place
Blocked filter
Incrustation on the float
Vacuum adjustment knob completely open
Blocked protection filter
Connecting tubing to the filter and device
clogged, bent or disconnected
Cannister lid overflow valve closed or
blocked
Dirty, blocked or damaged pump
16
SOLUTION
- Correctly insert the power cable into the
sockets
- Correctly position the collection cannister
lid
- Correctly position the gasket on the lid
- Replace the filter
- Remove the cannister lid and safety device
and extract the float. Continue by carrying
out the cleaning procedures as described in
the paragraph, "CLEANING, SANITISATION,
DISINFECTION, STERILISATION"
- Fully close the adjustment knob and check
the vacuum power
- Replace the filter
- Check the condition of the tubing, replace
it if blocked and correctly connect it as per
the "ASSEMBLY DIAGRAM" of this manual
- Unblock the overflow valve, keeping the
device upright
- Take the device to your trusted dealer or
an authorised FLAEM service centre