30
Frequency track B
0 * 10 kHz
1
25 Hz
2
3 Hz
Input logic
0 * PNP (operating point at 11 V)
1
NPN (operating point at 11 V)
2
PNP (operating point at 5 V)
3
NPN (operating point at 5 V, also for Namur but without ex-proof
4
PNP (operating point at 2.5 V)
5
NPN (operating point at 2.5 V)
Reset main counter
0 * Externally static (>30 ms)
1
Externally diff. (approx. 1 ms)
2
Externally diff. active in plus direction only (>30 ms)
3
Externally diff. active in minus direction only (>30 ms)
Reset key function enable
0 * Reset key active
1
Reset key disabled
Function, signal input 9 (stop / hold)
0 * Stop (approx. 1 ms)
1
Hold (approx. 1 ms)
2
Keylock
3
Print (>30 ms). Units with interface only.
Output signal assignment
0 * With 0-Signal
1
With start count value
Output signal logic
0 * Output as normally open contact
1
Outpu as normally closed
Output time (see 5.4 Output modes)
0 * Wiper function (pulse signal time in line 35)
1
Latch
2
Min. wiper signal at 0 / SC
Pulse signal time
0 * 0.25 s
0.01 to 99.99 s
Code
0 * Code not active
1 to 9999
Baud rate
0 * 4800 baud
1
2400 baud
2
1200 baud
3
600 baud