8) INFORMATION: The double casting process is used for lamination.
Cast the distal section of the prosthetic socket (2/3 of the length) with Orthocryl lamination
resin and allow it to cure.
9) Cut a piece of Perlon stockinette to twice the length of the plaster positive and pull on one
half.
10) Twist the Perlon stockinette on the distal end and pull on the other half.
11) Cast the prosthetic socket with Orthocryl soft and allow it to cure.
5.4 Static Alignment
+ = anterior placement / – = posterior placement (in relation to the load line)
Item
see fig. 10
Required equipment:
743L100 L.A.S.A.R. Posture
   
To determine the sagittal load line
Prosthetic foot (with shoe) on force measurement plate (sufficient load: > 35% body weight)
Other foot (with shoe) on height compensation plate
The tips of the shoes are in one line
   
Optimise the static alignment solely by changing the plantar flexion. Adjustment is performed only
with the anterior and posterior set screws of the tube adapter on the prosthetic foot.
a–p positioning of the alignment reference point
load line
-35 mm
m–l positioning, middle of the prosthetic foot to the frontal load line
0 to 20 mm medial
m–l positioning, centre of the prosthetic knee joint to the frontal load line
0 to 20 mm medial
m–l positioning of the prosthetic socket:
Positioning of the anterior superior iliac spine (Spina iliaca anterior superior)
line
:
0 to 20 mm lateral
5.5 Dynamic Trial Fitting
CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
Only adapt the settings to the patient gradually.
Explain the effects of the adjustments on the use of the prosthesis to the patient.
During dynamic trial fitting, the alignment and settings of the prosthesis are checked and adapted
for optimum walking according to the needs and abilities of the patient.
The patient has to learn the safe use of the prosthesis through intensive training.
The following subsections describe the adjustment possibilities for adapting the product to the
patient.
The list below provides an overview of the sequence for working through the subsections:
Checking the Factory Settings
Setting Flexion Damping
Setting Extension Damping
Static alignment process
:
, position the patient on the L.A.S.A.R. Posture as follows:
(upper, anterior rotation axis) to the sagittal
:
:
to the frontal load