Appendix
Nominal transmitter power
(W)
0.01
0.1
10
100
52/60
Downloaded from
www.Manualslib.com
8.1.4
Wellsystem Medical_Plus has been designed for use in an electromagnetic
environment in which the radiated HF interference is controlled. The operator
of the massage device can help to reduce electromagnetic interference by
maintaining a safe distance between the portable and mobile HF communica-
tion devices (transmitters) and the device itself as recommended below with
regard to the communication device's maximum output power.
150 kHz to 80 MHz
0.12
0.38
1
1.2
3.8
12
The recommended safe distance (d) in metres (m) can be determined by
using the equation that applies to the transmitters frequency in the case of
transmitters where their nominal power is not listed above, whereby 'P' is the
nominal transmitter power in Watts (W) according to the details supplied by
the manufacturer of the transmitter.
Comment 1: The safe distance for higher frequency ranges applies to 80 MHz
and 800 MHz.
Comment 2: The propagation of electromagnetic waves is affected by
absorption and reflexes caused by buildings, objects and people.
manuals search engine
Table 206: Recommended safe distances
between portable and mobile HF
communications equipment and
Wellsystem Medical_Plus
Safe distance with regard to the transmitting frequency
(m)
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
0.23
0.73
2.3
7.3
23