Herunterladen Inhalt Inhalt Diese Seite drucken

Rated Data; Special Conditions - Mettler Toledo O2 ISM Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für O2 ISM:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

14
2.8.2

Rated data

For rated data please observe section 2.7.2.
2.8.3

Special conditions

In type of protection intrinsically safe apparatus, the analytical probes
(oxygen ISM sensor simulators) equipment is designated with the
following specific conditions of use.
1. The oxygen ISM sensor simulators shall be installed in compli-
ance with the enclosure, mounting, spacing and segregation
requirements of the ultimate application, including a tool remov-
able cover.
2. Process temperature no greater than +130 ºC.
3. Potential Electrostatic Charging Hazard – To prevent the risk of
electrostatic sparking, the non-metallic surface should only be
cleaned with a damp cloth.
O
ISM Qualification Kit
2
52 300 440 E
© 12/18 Mettler-Toledo GmbH, Greifensee
O
ISM Qualification Kit
2
Printed in Switzerland

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis