2.8.2
Rated data
For rated data please observe section 2.7.2.
2.8.3
Special conditions
In type of protection intrinsically safe apparatus, the analytical probes
(oxygen ISM sensor simulators) equipment is designated with the
following specific conditions of use.
1. The oxygen ISM sensor simulators shall be installed in compli-
ance with the enclosure, mounting, spacing and segregation
requirements of the ultimate application, including a tool remov-
able cover.
2. Process temperature no greater than +130 ºC.
3. Potential Electrostatic Charging Hazard – To prevent the risk of
electrostatic sparking, the non-metallic surface should only be
cleaned with a damp cloth.