Herunterladen Inhalt Inhalt Diese Seite drucken

NSK Varios 570 Bedienungsanleitung Seite 10

Multi function ultrasonic scaler
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
How to use the Tip Card
1) Place the neck of the Tip in the cut out.
2) Check wear of the Tip.
3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we recommend
to replace a Tip when the Tip meets the yellow line (wear of 1mm) to guarantee safe and effective use.
Tip Card
The Tip Card can be used to check the following
tips : G1, G4, G6, G8, G9, P1/P1D, P10, and P20
CAUTION
Tips are consumables. The efficiency of dental scaling
decreases approximately 25% when the top of the Tip
wears 1 mm and approximately 50% when it wears 2 mm.
In addition, the vibration condition changes owing to the
wear, which may damage a patient's tooth surface. Check
the Tip wear condition with the Tip Card periodically, and
replace the Tip with a new one in good time.
9. How to Use Tip Cover S (Option)
Grip the Tip Cover S and insert it to the Tip.
To remove, grip the Tip Cover S and the handpiece & pull.
(Fig. 13)
The Tip Cover S is not designed for use as a Tip changing
tool.
CAUTION
Carefully insert the Tip
into the Tip Cover S. Avoid
injuring the fingers.
Neck
Handpiece
Fig.11
Slit
Green: No wear - Tip is OK
Tip replacement is not necessary.
Yellow: Wear of 1mm - Tip is showing some wear
Tip replacement is recommended.
Red: Wear of 2mm - Tip is badly worn
Tip replacement is necessary.
1mm
25%
Decrease
Slit
Tip Cover
9
2mm
50%
Decrease
Efficiency
Fig.12
Tip
Tip-Handpiece Joint
Fig.13

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis