Bubble CPAP System - BC151/BC161 User Instructions
MR850: 37°
9. Set the Humidifier
Ensure there is air flow present before turning
on the humidifier.
• Turn on the humidifier.
- If using the MR850, ensure it is on intubation
mode.
- If using the MR730, set the temperature to
40°, -3.
• Refer to MR850 or MR730 user instructions for
further information.
10
MR730: 40°, -3
BC161
10. Attach Interface to Infant
Connect the interface to the infant
using the instructions provided with the
interface.
9
8
7
6
Checks During Operation
• Regularly observe that the water is feeding into
the humidification chamber.
• Should the water level exceed the maximum
level marked on the humidification chamber,
the chamber should be replaced.
• Check all connections are tight before use and
after any adjustment.
• Ensure air flow is present at all times. If air flow
is interrupted, turn off the humidifier.
• Regularly observe the circuit for condensate.
Drain as required in order to maintain set CPAP.
• Regularly observe the CPAP generator for
bubbling. If bubbling is not observed, check for
and minimize air leaks in the system and at the
BC151
patient. If air leaks have been minimized, air
flow may be increased to achieve continuous
bubbling.
• Regularly observe the water level in the CPAP
generator and overflow container. Refill the
CPAP generator if the water level drops below
the minimum water level line. Check and empty
the overflow container once every 8 hours.
• Monitor Patient Oxygen levels.
• Always use pressure monitoring to verify that
the patient is receiving the prescribed CPAP
level.