Bubble CPAP System - BC151/BC161 User Instructions
1 L/min
5. Leak Test
Check all connections are tight before
use.
• Set input flow rate to 1 L/min.
• Observe the CPAP generator. Gentle,
audible bubbling is acceptable, no
bubbling means unacceptable leakage.
• If no bubbling is observed, check the
entire system.
Remove the correct port cap from the
pressure manifold before connecting any
monitoring device.
6-8 L/min
(<15 L/min)
6. Set Flow Rate
• Adjust the flow rate to the prescribed
input flow rate.
• Recommended 6-8 L/min.
• Allowable flow range: 4-15 L/min.
Pressure-Flow Table:
FPH Bubble CPAP Generator used with FPH Infant Interface, Typical mean pressure values (cmH
3
4
5
6
Flow
(L/min)
3.1
4.1
5.1
6.1
7.1
4
3.2
4.2
5.3
6.3
7.3
5
3.4
4.4
5.4
6.4
7.4
6
3.6
4.6
5.6
6.6
7.6
7
3.8
4.8
5.8
6.8
7.8
8
4.0
5.0
6.0
7.0
7.9
9
1 0
9
8
7
6
7. Set CPAP Level
• The number on the probe above the lid
indicates the CPAP pressure in cmH
• Set the CPAP probe at the prescribed
level (3-10 cmH
O).
2
Probe setting (cmH
7
8
9
10
Flow
(L/min)
8.1
9.1
10.0
10
8.3
9.2
10.1
11
8.4
9.3
10.2
12
8.5
9.5
10.4
13
8.7
9.6
10.5
14
8.9
9.8
10.6
15
8. Connect Bubble CPAP Circuit to Infant
Interface
Remove the flow test elbow and connect
O.
2
the circuits to the Infant Interface using
the instructions provided with the
interface.
• BC161 connects to F&P Infant Interface.
• BC151 connects to short binasal prongs.
(e.g. Hudson prongs)
2
O)
2
3
4
5
6
7
4.3
5.3
6.2
7.2
8.1
4.6
5.6
6.5
7.4
8.4
4.9
5.8
6.8
7.7
8.5
5.2
6.1
7.0
7.9
8.8
5.5
6.4
7.3
8.2
9.0
5.8
6.7
7.5
8.4
9.2
O) with no leak
8
9
10
9.0
9.9
10.8
9.2
10.1
10.9
9.4
10.2
11.1
9.6
10.4
11.3
9.8
10.6
11.4
10.0
10.8
11.6