Use this menu to set the Legal for Trade status.
On = operation complies with Legal Metrology regulations.
When Approved Mode is set to On, the menu settings are affected as follows:
Calibration Menu:
For InCal models, only Internal Calibration is available. All other functions are hidden.
Balance Setup Menu:
Filter Level is locked at the current setting.
Auto Zero Tracking is limited to 0.5 Division and Off. The selected setting is locked.
Auto Tare is locked at current setting.
Graduations are forced to 1 Division and the menu item is hidden.
Communication Menu (Communication->Print Settings->Print Output):
Stable Weight Only is locked On.
Numeric Value Only is locked Off.
Communication Menu (Communication->Print Settings->Auto Print):
Auto print mode selections are limited to Off, On Stability, and Interval. Continuous is not available.
Lockout Menu:
Menu is hidden
The security switch located at the rear of the balance must be in the locked position to set Approved Mode to
On. The security switch must be in the unlocked position to set Approved Mode to Off. See Section 6.
Enter this sub-menu to activate the desired units of measure.
PX balances provide a choice of 21 units, which are all set On by default.
Due to national laws, the balance may not include some of the units of measure listed.
g
kg
t
mg
ug
ct
N
lb
oz
ozt
GN
1.800.561.8187
Gram
Kilogram
Ton
Milligram
Microgram
Carat
Newton
Pound
Ounce
Troy ounce
Grain
www.
dwt
mo
msg
tl H
tl S
tl T
tcl
tola
baht
lboz
information@itm.com
.com
Pennyweight
Momme
Mesghal
HK tael
SG tael
TW tael
Tical
Tola
Baht
lb:oz

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Px85Px125dPx225dPx84Px84/ePx124 ... Alle anzeigen

Inhaltsverzeichnis