Herunterladen Inhalt Inhalt Diese Seite drucken

Disposal; Specifications - Welch Allyn KleenSpec 78812 Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 65
Wet Steam Autoclave
Expendable Illuminator (REF 78600)
1.
Clean expendable illuminator as instructed under "Cleaning - Soak".
2.
Place the expendable illuminator in the autoclave chamber to allow best exposure to
steam. Autoclave chamber parameters are as follows:
•Type:
•Exposure Temperature:
•Exposure Time:
•Exhaust Time:

Disposal

There are no harmful materials associated with the Illumination System.
Device is considered a biohazard after use. Follow good hospital practices and local
ordinances in the disposal of the illuminator.
There are no risks associated with the disposal of this device after its useful life if it is
cleaned and/or disinfected per the instructions in this manual.

Specifications

Classification:
Input:
Output:
Category:
Physical Specifications:
Illuminator:
Power Supply:
Illumination:
Lamp Life:
Lamp Voltage:
Lamp:
IPX Rating:
Illuminator:
Power Supply:
Environment:
Operating:
Transport/Storage:
Intermittent Operation:
Vag Spec and Illuminator
Expendable Illuminator only (78600)
°
132
Pre-vacuum (Hi-VAC) Steam Cycle
132
°
C to 135
°
C (270
°
F to 275
3 - 5 minutes
Fast
Class II, Type BF Applied Part
100V~ - 240V~, 50-60Hz, 200mA
Not AP/APG Equipment
Welch Allyn part no. 08800
+10° C to + 35° C (50° F to 95° F), 95% RH
500 hPa to 1060 hPa
-20° C to +49° C (-2° F to 120° F), 95% RH,
500 hPa to 1060 hPa
10 minutes on max; 5 minutes off min
8 minutes on max; 5 minutes off min
1-6
°
F)
4.7 V
, 1300mA
91.4 cm (36 in)
183 cm (72 in)
100 hours
4.6 V
IPX7
IPXØ

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Kleenspec 78814Kleenspec 78816

Inhaltsverzeichnis