Herunterladen Inhalt Inhalt Diese Seite drucken

Fluke 917X Series Erste Schritte Seite 27

Metrologie-blockkalibratoren
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 139
917X Series Metrology Wells
Getting Started
2. To enable "Probe 1," use the 4 or 5 button to turn the reference probe "On." Press
"ENTER" to accept the selection.
3.
Press the "Setup" soft key (Main Menu: R
setup menu, select "Probe 1" and press "ENTER".
4.
Select the calibration type for the p
"ENTER" to accept the calibration type.
5.
Use the numeric keys to enter the serial number and cal
Press "ENTER" after each parameter is entered to accept the new parameter value.
Calibration coefficients can be found on a
your probe. If no coefficients can be found, contact the manufacturer or vendor of the
probe for assistance. If the probe is out of calibration, Fluke's Hart Scientific Division
offers calibration services. Contact an Authorized Service Center for assistance on
obtaining a calibration for the probe.
6.
After the reference probe has been enabled, the focus of the temperature display can be
set. The focus is disabled if the reference probe is disabled. From the reference men
press the "Focus" soft key (Main Men
select reference focus. Press "ENTER" to accept the selection. The largest temperature o
the display will now be the reference temperature.
18
EF MENU|SETUP). In the reference probe
robe (ITS-90 or CVD) using the 4 or 5 keys. Press
ibration coefficients of the probe.
report of calibration that was shipped with
u: REF MENU|FOCUS). Use the 4 or 5 key to
u,
n

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

9171917091729173Hart scientific 917 serieHart scientific 9170 ... Alle anzeigen

Inhaltsverzeichnis