Setup and Assembly

Note
When selecting a setup location for the L9000, consult the
"Electromagnetic Compatibility" section included in this manual
to determine the best location.
To set up the L9000, make the following connections:
• Connect the AC power cable
• Connect the light cable
Caution
To ensure proper cooling, position the L9000 on its feet.
Do not operate the unit upside down or on its side.
W
B
ARNING
ENDOSCOPE TO ENSURE THERE ARE NO ROUGH SURFACES
SHARP EDGES
O
SPECIFIED FOR USE WITH THE LIGHT SOURCE
INCOMPATIBLE EQUIPMENT MAY CAUSE UNEXPECTED RESULTS

Connecting the AC Power Cable

1.
Plug in the AC power cord to the AC Inlet on the rear console panel.
2.
Plug in the other end of the AC cord to a hospital-grade outlet.
W
W
ARNING
ELECTRICAL EQUIPMENT
T
O MINIMIZE LEAKAGE CURRENT THAT MAY TRAVEL TO THE
PATIENT OR USER
ONLY WITH OTHER
SYSTEMS ARE INSTALLED ACCORDING TO THE REQUIREMENTS OF
IEC 60601-1-1.

Connecting the Light Cable

W
U
ARNING
L9000
Note
The L9000 Light Source is compatible with all Stryker Light
Cables.
10
,
EFORE EACH USE
,
OR PROTRUSIONS
NLY CONNECT ITEMS TO THE LIGHT SOURCE THAT HAVE BEEN
L9000
HEN THE
IS INTERCONNECTED WITH OTHER MEDICAL
,
ANY
SE ONLY NONCONDUCTIVE FIBEROPTIC CABLES WITH THE
TO MAINTAIN ELECTRICAL ISOLATION
CHECK THE OUTER SURFACE OF THE
.
,
LEAKAGE CURRENTS MAY BE ADDITIVE
T
CF
YPE
APPLIED PART SHOULD BE USED
T
CF
YPE
APPLIED PARTS
,
. C
ONNECTING
. E
NSURE ALL
.
.
.