Selecting the Operation Mode

The L9000 has two operation modes, Activate and Standby.
• ACTIVATE mode: The Activate mode is used during normal
operation. It enables light output to be controlled by the brightness
controls on the front console panel.
• STANDBY mode: The Standby mode is used when the L9000 is
powered on, but not in use. It reduces the light output to a minimum,
thereby reducing the heat generated at the tip of the light cable or
scope when the L9000 is not being used.
Press the Activate/Standby button on the LCD screen to toggle between the two
operational modes.
W
T
ARNING
O HELP PREVENT BURNS TO THE PATIENT
OBJECTS AND POSSIBLE FIRE
S
TANDBY MODE WHENEVER THE ENDOSCOPE IS REMOVED FROM
THE LIGHT CABLE
CABLE ADAPTER
MINUTES TO COOL OFF AFTER BEING PLACED IN STANDBY MODE
AND THEREFORE MAY STILL RESULT IN FIRE AND
THE PATIENT
PROPERLY
THE PATIENT
EVEN WHEN THE DEVICE IS IN STANDBY MODE
BRIGHTNESS
STANDBY
. T
HE SCOPE TIP
,
AND LIGHT CABLE TIP WILL TAKE SEVERAL
,
USER OR INANIMATE OBJECTS IF NOT USED
. D
O NOT PLACE THE SCOPE OR THE LIGHT CABLE ON
,
ON THE DRAPES
,
USER OR INANIMATE
,
ALWAYS PUT THE
,
SCOPE LIGHT POST
,
OR OTHER FLAMMABLE MATERIAL
L9000
IN
,
LIGHT
/
OR BURNS TO
.
13
,