Models 947, 948 and 960 Leak Detectors Operations and Service Manual
Table D-2 Optional Discrete I/O Parallel Interface (J120) (Continued)
6
Parallel Enable
7
Zero
9
Auto Cal
10
Decrement
11
Increment
12
Auto Range
14
Hold
15
Filament Fault
16
Not Ready
18
CAL OK
19
Zeroing
21
Calibrating
23
Ready
27, 28
+24 VDC for Output Sourcing
D-2
Active High
200 msec high pulse
200 msec high pulse
200 msec high pulse
200 msec high pulse
Active High
Active High
Active High
Active High
Active High
Active High
Active High
Enables rear panel
PLC Interface.
Begins Auto Zero routine.
Begins Auto Calibrate.
Decrements the exponent
when in manual ranging.
Increments the exponent
when in manual ranging.
Automatically changes the
exponent to keep the reading
in range.
Not used.
Active when the selected
ion source filament is
non-functional.
Not ready to run – possible fault. Output
Active when last
CALIBRATE successful.
Active when Auto Zeroing.
Active when Auto Cal.
Active when system is operating
(inverse of "Not Ready").
Input
Input
Input
Input
Input
Input
Input
Output
Output
Output
Output
Output
Output