Models 947, 948 and 960 Leak Detectors Operations and Service Manual
[SHIFT]
[SHIFT]
[SHIFT]
[SHIFT]
[SHIFT]
A-6
Sets a permanent new value for Emission Current. Enter XXXX
INIT
(the desired value), followed by the SHIFT key, then the
EMIS
PUT
function key.
ZERO
Initiates an atmosphere calibration of the Test Port TC.
TP TC
ATM
ZERO
Initiates a zero calibration of the Test Port TC.
TP TC
ATM
ZERO
Initiates an atmosphere calibration of the System TC.
SY TC
ATM
NOTE
CAUTION
ZERO
Initiates a zero calibration of the System TC.
SY TC
ATM
SET
Displays the value of the Auto Sequencer Test Time.
TST TM
SHOW
Sets the value of the Auto Sequencer Test Time. Enter XXX
SET
(the desired value), followed by the SHIFT key, then the
TST TM
SHOW
function key.
SET
Displays the value of the Auto Sequencer Roughing Time.
RUF TM
SHOW
Sets the value of the Auto Sequencer Roughing Time. Enter XXX
SET
(the desired value), followed by the SHIFT key, then the
RUF TM
SHOW
function key.
ON
Turns off the Auto Sequencer function.
A SEQ
OFF
System TC being at atmosphere is not a normal state for
the leak detector. The turbo pump must be turned off,
and the spectrometer tube must be vented by operating
the valves or by opening the vent screw on the turbo
pump. This is best done before the pump gets up to
speed just after turning the leak detector on.
You must take care to avoid damage to the vacuum
system and the spectrometer tube. Follow instructions
carefully.