the rear of the unit.
• The fuse compartment is located directly above to the power cord receptacle. To access, take a
small, flat-headed screwdriver to the tabs on the fuse holder. Apply slight pressure to release
the fuse holder
(Figure
• Gently remove the fuses from the fuse holder.
• Examine the fuses
(Figure
current. If fuse appears to be in good condition return the unit to your dealer or Coltène/
Whaledent.
• Plug the power cord into the power cord receptacle on the rear of the unit. Then plug the power
cord into the wall receptacle.

13. Technical Details

Environmental Conditions:
• Intended for indoor use.
• Maximum altitude 2000m.
• Recommended Storage and transport temperature:
• Temperature during operation:
• Relative humidity:
• Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage.
• IEC -60664-1 Pollution degree 2.
Electrical Information:
• Power Supply Input: 100V – 240V AC / 50 ~ 60 Hz
• Transducers: 2
• Operating Frequency: 45 kHz - 47.5 kHz
• Fuse Specifications: 2 Fuses FST, 5 × 20 mm, Slow Blow, 250VAC, 5A, IEC 60127-2 Sheet 3
9).
10). If blown, replace with a fuse that matches the unit's voltage and
13