Herunterladen Inhalt Inhalt Diese Seite drucken

Benning it100 Bedienungsanleitung Seite 16

Vorschau ausblenden Andere Handbücher für it100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
5.1.15 A fter switch on the battery symbol appears at the top right in the display
for about 5 seconds (
tery capacity is indicated by four segments. The battery symbol persists
if only one segment or none is displayed. The battery symbol appears
during operation and in the case of further falling voltage only when the
switch is actuated (see also Section 7.5 and Section 8.6).
Replace­­ the­­ batteries­­ immediately­­ with­­ fresh­­ ones­­ when­­ all­­
segments­­ of­­ the­­ battery­­ symbol­­ are­­ extinguished,­­ to­­ avoid­­
endangering­­persons­­by­­false­­measurements.
5.1.16 W ith fresh batteries the BENNING IT 100 can make about:
- 6000 insulation resistance measurements (250 V) or
- 5000 insulation resistance measurements (500 V) or
- 3w500 insulation resistance measurements (1000 V)
(according to EN 61557-2), or
- 4000 low resistance measurements (according to EN 61557-4).
5.1.17 Dimensions of the instrument (L x W x H) = 235 x 100 x 55 mm
Appliance weight:
6.­­­­ Environmental­­conditions
- T he BENNING IT 100 is intended for making measurements in dry environ-
ments.
- Maximum barometric elevation for making measurements: 2000 m.
- O vervoltage category IEC 60664/ IEC 61010, 600 V category III, 1000 V
category II
- Contamination degree 2
- Protective system IP 40 (EN 60529)
I P 40 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 1 mm, (4 - first index). No protection
against water, (0 - second index).
- E lectromagnetic compatibility (EMC): Interference immunity and emitted
interference conforming to EN 61326-1
- O perating temperature and relative humidity:
For operating temperature from 0 °C to 31 °C: relative humidity less than 95 %
F or operating temperature from 31 °C to 40 °C: relative humidity less than 75 %
F or operating temperature from 40 °C to 50 °C: relative humidity less than 45 %
- S torage temperature: The BENNING IT 100 can be stored at any tempera-
ture in the range from - 25 °C to + 65 °C (relative humidity up to 90 %). The
batteries should be taken out of the instrument for storage.
7.­­­­ Electrical­­specifications
Note: The measuring accuracy is specified as the sum of
- a relative fraction of the measured value and
- a number of digits (counting steps of the least significant digit).
The specified measuring accuracy is valid for temperatures in the range from 10
°C to 30 °C and relative humidity less than 90 %.
7.1­­­­ I nsulation­­resistance­­measuring­­ranges
(switch setting MΩ, 250 V/ 500 V/ 1000 V)
Measuring range
 MΩ
0 MΩ
00 MΩ
000 MΩ
The indicated measuring accuracy is specified for the following insulation resist-
ance values and depends on the test voltage:
Test voltage
Analog bargraph display:
Accuracy:
11/ 2006
450 g without batteries
590 g with batteries
Resolution
0,0 MΩ
0, MΩ
 MΩ
10 MΩ
50 V
500 V
000 V
0 to 1 G and infinity
10 %
BENNING IT 100
,
,
,
,
Accuracy
 % ±  digits
 % ±  digits
 % ±  digits
6 % ±  digits
). The still available bat-
Overload protection
000 V AC/ DC
000 V AC/ DC
000 V AC/ DC
000 V AC/ DC
range
0,5 MΩ - 1000 MΩ
0,5 MΩ - 2000 MΩ
 MΩ - 2000 MΩ


Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis