Configuration and Adjustment
The configuration and adjustment of the EE100Ex is only permitted outside the hazard area.
Connection of CH1 or CH2 is NOT allowed for communication, see connection example below:
USB:
PC
Fig. 5
Adjustment and configuration
4
Maintenance
Operation and maintenance in explosion hazard area may be performed only by trained specialist
personnel authorised to do so by the system operator.
The standards EN60079-17 or IEC60079-17, EN60079-19 or IEC60079-19 as well as all relevant natio-
nal regulations shall be observed for maintenance and repair work in explosion hazard areas.
When employed in dusty, polluted environment:
• The filter cap shall be replaced once in a while with an E+E original one. A polluted filter cap causes
longer response time of the device.
• If needed, the sensing head can be cleaned. For cleaning instructions
please see www.epluse.com/ee100ex.
The cleaning of the sensing element may be only performed in a non-hazardous area.
EE100Ex Safety Guide
Connection cable - HA011068
Configuration adapter - EE-PCA
DO NOT CONNECT
9