CAUTION
If the chuck is not regularly cleaned the chuck grip may be weakened and the bur may be accidentally released while in operation.
NSK Care3 Plus
When using NSK Care3 Plus automatic handpiece cleaning and lubrication system, refer to the Care3 System instructions.
7-4 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.