Troubleshooting/FAQs

4.2 Alarm Letter Codes and Descriptions

Message
Low Signal - Press <ENT>
Detection Fault
4.2
Alarm Letter Codes and Descriptions
Letter Code
S
0
E
R
F
A
M
K
I
P
The display shown below indicates where the Alarm Codes appear on the LCD display
screen.
30
During the Initial Makeup the flow meter decides that the level of the receive signal is
insufficient for proper operation. Some reasons for low signal are:
Invoking [Install completed?] on an empty pipe.
Coupling compound insufficient; not applied or evaporated.
A disconnected or broken transducer cable.
The pipe needs to be conditioned at the mounting location.
Flush out large air bubbles.
The Xdcr cables are defective or not connected to the correct channel.
The Set Empty routine performed when pipe was NOT actually empty.
If you locate and correct the improper condition immediately, press <ENT> to resume the
installation procedure. Otherwise, press the <Left Arrow> to abort the installation and
conduct a thorough investigation.
If it appears that the flow meter cannot complete an Initial Makeup it means that the pipe
and/or liquid conditions do not permit a receive signal that meets the flow detection
standards. The system will not operate. Attempt to improve operating conditions by
reinstalling the transducers at a different spacing offset, or even at a different location on
the pipe.
Switching from Reflect to Direct Mount may solve the problem. However, operation may
not be possible if there is poor liquid or pipe wall sonic conductivity.
Alarm
Spacing
Transducer spacing may need readjustment.
ZeroMatic
ZeroMatic signal fault.
Empty
Pipe is empty.
Rate
Flow above High setting or below Low setting.
Fault
Three continuous seconds without new data update.
Aeration
Current aeration percentage exceeds the alarm set point.
Memory
Last valid reading for a selected interval during Fault condition.
Makeup
In-Process Makeup occurred.
Interface
Liquid Vs exceeds interface alarm set point.
Pig
Pig passage detected (optional).
Description
Description
SITRANS FUS1020 IP65 NEMA 4 Quick Start