2.
Connect the video output.
• The rear panel provides three analog and three digital-video outputs,
which can be used together or independently:
Output
Output
Type
Analog
Composite
*S-VHS 1
*S-VHS 2
Digital
**DVI-I 1
**DVI-I 2
DVI over optical
FIBER
**On some monitors, S-VHS inputs may be labeled Y/C.
**The DVI connectors can also output analog SXGA signals through a DVI-I to VGA adapter.
• Use the cables and outputs described above to connect the 1188HD to
other operating-room equipment. Wiring Diagrams 1-3 on the
following pages describe typical set-ups.
• If desired, connect any remote outputs using the remote cables
supplied with the 1188HD Camera. (See Wiring Diagram 2.) Devices
connected to the remote outputs of the 1188HD Camera can be
operated using the P buttons on the camera head and/or console. See
the "Operation Instructions" section of this manual for details.
• If desired, connect the SIDNE® interface as well. (See Wiring
Diagram 2.)
Warning
When the 1188HD Camera is used with other equipment,
leakage currents may be additive. Ensure that all systems are
installed according to the requirements of IEC 60601-1-1.
Warning
Do not touch the internal pin of the VIDEO-OUT BNC jack
and the patient simultaneously.
Cable
Connector
Composite
BNC
(push-and-turn connectors)
S-VHS
4 pin Mini-Din
(push-only connectors)
S-VHS
4 pin Mini-Din
(push-only connectors)
DVI
29-pin (push-only connectors, with two
tightening knobs)
DVI
29-pin (push-only connectors, with two
tightening knobs)
Fiber (×4)
Lucent connector fiber (×4) (push-only)