Cleaning: Manual
1. Brush
• Thoroughly brush the exterior of the device with a soft-bristled brush,
focusing on any mated or rough surfaces.
• Inject any lumen or mated surface a minimum of five times with at least
50mL of the detergent.
• Brush any lumens a minimum of five times from each end, using an
appropriate bottle brush.
• Brush any movable parts in their extreme open and closed positions.
2. Rinse
• Rinse the device with treated water at ambient temperature to remove all
detergent residue. Flush any lumens or mated surfaces a minimum of 5
times. Once all detergent residues have been removed, continue to rinse
for a minimum of 30 seconds.
• Drain excess water from the device and dry it using a clean cloth or
pressurized air.
• Visually inspect the device for cleanliness, paying close attention to hard-
to-reach areas. If visible soil remains, repeat steps 1 and 2.
3. Soak
• Prepare a non-enzymatic detergent, according to the manufacturer's
recommendations of 0.25 ounces/gallon tap water at 35 - 40°C.
• Fully immerse the device and inject any lumens and mated surfaces with
at least 50mL of the detergent.
• Soak the device for a minimum of 15 minutes.
4. Brush
• Thoroughly brush the exterior of the device using a soft-bristled brush.
• Inject the prepared detergent into any cannulae, lumens, or mated
surfaces a minimum of 5 times.
• Brush any lumens a minimum of 5 times from each end, using an
appropriate bottle brush.
• Actuate the device, brushing around any movable parts in all extreme
positions.
5. Rinse
• Thoroughly rinse the device with treated water until all detergent residue
is removed. Flush any lumens or crevices 5 times. After the detergent
residue is removed, continue rinsing for a minimum of 30 seconds.
• Drain the excess water from the device and dry it with a clean cloth or
pressurized air.
EN-31