Cautions
• Always install the soaking cap prior to processing the camera.
Failure to properly tighten the soaking cap will corrode the
connector pins and void the warranty.
• Inspect the camera cable for cuts and breaks before soaking in any
fluid. Return any damaged camera to Stryker for service.
• Never soak the camera in the same tray with sharp instruments.
• Do not use brushes or pads with metal or abrasive tips during
manual cleaning, as permanent scoring or damage could result.
• To minimize galvanic corrosion, avoid soaking dissimilar metals in
close proximity.
• Only camera heads marked
sterilization. Autoclaving camera heads that do not bear this
marking will result in product damage.
• Allow the device to air cool following steam sterilization. Rapid
cooling or "quenching" in a liquid will damage the device and void
the warranty.
• Allow the camera head to cool before connecting it to the console.
Connecting the camera head while it is still hot may result in
system error.
Limitations on Reprocessing
• Do not cross-sterilize the device. Using multiple sterilization
methods may significantly reduce the performance of the device.
• Do not leave the device in solutions longer than necessary. This
may accelerate normal product aging.
• Proper processing has a minimal effect on this device. End of life is
normally determined by wear and damage due to use.
• Damage incurred by improper processing will not be covered by the
warranty.
can withstand steam
AUTOCLAVE
EN-29