8
1.2.3

Special Conditions

– The relationship between the maximum permissible ambient or
media temperature and temperature class, for category 1G appli-
cations, zone 0, is shown in the following table:
Temperature class
T 6
T 5
T 4
T 3
– The relationship between the maximum permissible ambient or
media temperature and temperature class, for category 1D appli-
cations, zone 20, is shown in the following table:
Temperature class
T69°C
T81°C
T109°C
T161°C
– The capacitance and inductance of the connecting cable has to
be considered.
– The O2 ISM sensor plug-in heads can be used in / with the fittings
InFit 76*-*** or InTrac7**-***, or in / with other suitable fittings in
potentially explosive areas.
– The independent fitting used for installation of O
in head must be conductively connected to the equipotential
bonding system.
O
ISM Qualification Kit
2
52 300 440
Max. ambient or
media temperature
68 °C
80 °C
108 °C
160 °C
Max. ambient or
media temperature
68 °C
80 °C
108 °C
160 °C
© 05/15 Mettler-Toledo AG, Greifensee
O
ISM Qualification Kit
2
ISM sensor plug-
2
Printed in Switzerland