Mettler Toledo O2 ISM Bedienungsanleitung Seite 11

Qualifizierkit
Vorschau ausblenden Andere Handbücher für O2 ISM:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

O
ISM Qualification Kit
2
Problem solving: If any of the tests described above failed, you
should change the connecting cable and repeat the test. If the test
still shows the same result, the transmitter or the cable is defective.
If you are using a transmitter with current outputs, so ensure the out-
puts are correctly wired.
You must never do a calibration (the simulators are read-only).
You must never increment autoclave cycles (always set "NO" to cy-
cle counter), this is only needed if autoclave has been set to a value
higher than 0.
Different transmitters may display different values and errors.
Out of range behavior is also different depending on the transmitter
in use thus different values/readings may occur.
Set the transmitter input filter to off (only for SIM O
ppb TGL and SIM O
Do not use the O
on the label of each plug-in head. Please send them back to
METTLER TOLEDO for recertification.
© 05/15 Mettler-Toledo AG, Greifensee
Printed in Switzerland
Trace TGL).
2
ISM plug-in heads after the expiry date shown
2
TGL, SIM O
2
O
ISM Qualification Kit
2
52 300 440
11