Procedure for setting the DIP switch or activating the proof test:
1.
Remove the cover clamp.
2.
Unscrew the housing cover together with the O-ring.
3.
If necessary, remove the display with the fitting kit from the electronics module.
4.
Configure the hardware write protection WRITE LOCK accordingly using the DIP switch.
In general, the following applies: switch to ON = function enabled, switch to OFF =
function disabled.
5.
If performing a SIL commissioning test and a proof test, make a device restart using the
button.
Once the hardware setting has been made, re-assemble the housing cover in the reverse
order.