10.WHAT DO I DO IF ...?

Fault
... no weight is dis-
played when scale is
loaded?
... 0.000 does not appear
before weighing?
... a segment lights up
continuously or not at
all?
...
is displayed?
... the display "bAtt"
appears?
... the display "StOP"
appears?
... the display "tEMP"
appears?
CAUTION!
Personal injury
If you notice any faults or deviations during the
function check which cannot be resolved with
the aid of the section entitled "What do I do if
..." in this document, you may not use the
device.
– Have the device repaired by seca service or
by an authorized service partner.
– Follow the section entitled "Servicing/
recalibration" in this document.
Cause/Remedy
The scale is not connected to the power supply.
- Check if the scale is switched on.
- Check whether the battery block is inserted
and that the plug contacts are properly con-
nected.
A load was placed on the scale before it was
switched on.
- Remove load.
- Switch scale off and back on again.
The relevant segment is faulty.
- Please notify maintenance service.
Battery block power is running down.
- Charge battery block.
Battery block discharged.
- Charge battery block.
The maximum load has been exceeded.
- Remove load.
The ambient temperature of the scale is too high
or too low.
- Set up scale in an ambient temperature
between +10 °C and +40 °C.
- Wait approx. 15 minutes until scale has
adjusted to ambient temperature.
English
What do I do if ...? • 77